Protocol: PK Bridging ELISA for Use with Anti-Risankizumab Antibodies
Protocol: PK Bridging ELISA for Use with Anti-Risankizumab Antibodies
Download PDF
Pharmacokinetic (PK) Bridging ELISA: For use with Anti-Risankizumab Monoclonal Antibodies catalog TZA058, TZA059P
This method provides a procedure for carrying out a PK ELISA with Anti-Risankizumab Antibodies TZA058 (capture antibody), and HRP-conjugated TZA059 / TZA059P (detection antibody), using risankizumab for the standard curve. The method should always be used in conjunction with product and batch specific information provided with each vial (see product datasheets). This protocol will need to be adjusted for use with different detection methods and immunoassay technology platforms.
View all of our anti-risankizumab antibodies
Reagents
- BSA (Sigma-Aldrich, A7906)
HISPEC Assay Diluent (BUF049)
Human Serum (Sigma-Aldrich, H4522)
LYNX Rapid HRP Antibody Conjugation Kit™ (LNK001P-LNK006P)
- PBS
- 136 mM NaCl
2.68 mM KCl
8.1 mM Na2HPO4
1.46 mM KH2PO4
- PBST
- PBS with 0.05% Tween 20 (Merck Millipore, 817072)
- SpyCatcher Reagents e.g. BiSpyCatcher2:HRP (TZC002P). Contact us to discuss alternative SpyCatcher options
- QuantaBlu Fluorogenic Peroxidase Substrate (Thermo Fisher Scientific, 15169)
Materials
- 384-well microtiter plate, black, square flat-bottom wells, for example, Black 384-Well Immuno Plates (Thermo Fisher Scientific, 460518)
- Fluorescence plate reader
- 96-well plates can be used instead of 384-well plates, (black, flat-bottom wells) for example, Black 96-Well Immuno Plates (Thermo Fisher Scientific, 437111). For the 96-well format, use 100 μl (instead of 20 μl) of antigen, antibodies, or substrate and 300 μl for the blocking step.
Method
-
Prepare the capture Anti-Risankizumab Antibody TZA058 (AbD59012ad) at 1 µg/ml in PBS. Coat the required number of wells of a 384-well microtiter plate with 20 µl per well of the prepared capture antibody, and incubate overnight at 4°C.
-
If using TZA059P (AbD58231pap) for detection proceed to step 3. If using TZA059 for detection, prepare the detection Anti-Risankizumab Antibody: Couple TZA059 (AbD58231ad) to a suitable SpyCatcher Reagent (e.g.TZC002P) using the appropriate coupling protocol.
-
Wash the microtiter plate five times (5x) with PBST.
-
Block the microtiter plate by adding 100 µl 5% BSA in PBST to each well, and then incubate for 1 hr at RT.
-
Wash the microtiter plate 5x with PBST.
-
For the standard curve, prepare a dilution series of risankizumab in 10% human serum in PBST in triplicate. Final concentration of risankizumab should cover the range from 0.002 ng/ml to 1,000 ng/ml. Include a zero risankizumab concentration as the background value.
-
Add 20 μl of risankizumab dilution per well (in triplicate for each standard recommended). Add 20 μl of each test sample to the other wells (in triplicate for each sample recommended). Incubate for 1 hr at RT.
-
Wash the microtiter plate 5x with PBST.
-
To each well, add 20 µl of either (Bi)SpyCatcher:HRP-coupled detection Anti-Risankizumab Antibody, TZA059 (AbD58231ad), or HRP-conjugated detection Anti-Risankizumab Antibody, TZA059P (AbD58231pap) at 0.5 µg/ml in HISPEC Assay Diluent. Incubate for 1 hr at RT.
-
Wash the microtiter plate 10x with PBST.
-
Add 20 µl QuantaBlu Fluorogenic Peroxidase Substrate to each well and measure the fluorescence after 30 min.