Below are a list of common problems encountered within ELISA together possible causes and subsequent actions / solutions:
-
No signal
-
Weak Signal
-
High Background Signal
-
Slow Color Development
For further help and expert advice, contact our scientist staffed technical support department here.
Problem 1: No signal
Possible causes and solutions for no signal:
a. Assay set up incorrectly or used incorrect reagents
-
Review protocol. Repeat assay using a positive control
b. Incorrect secondary antibody used
-
Retrace steps. Repeat assay using the correct secondary antibody
c. Not enough antibody used
-
Increase concentration of the primary and/or secondary antibody
d. Detection reagent too old or contaminated
-
Use fresh detection reagents
e. Antigen not coated properly
-
Try longer coating times, different coating buffers, or avidin plates with biotinylated antigen
f. Plate reader has the wrong settings
-
Check plate reader for wavelength, filters, gain etc.
g. Antibody stored at 4°C for several weeks or subjected to repeated freeze/thaw cycles
-
Use a fresh aliquot of antibody that has been stored at -20°C or below
Problem 2: Weak signal
Possible causes and solutions for weak signal:
a. Insufficient amount of antigen was coated to microtiter plate
-
Use more antigen for coating or very coating buffer
b. Not enough antibody used
-
Increase concentration of the primary and/or secondary antibody. Optimize antibody concentrations for your assay
c. Detection reagent too old, contaminated, or used at the wrong pH
-
Use fresh detection reagents at the correct pH
d. Detection reagent too dilute
-
Use a higher concentration of detection reagent
e. Plate reader has the wrong settings
-
Check plate reader for wavelength, filters, gain etc
f. Incubation temperature too low
-
Optimize the incubation temperature for your assay. Reagents should be at room temperature before beginning the assay
Problem 3: High background signal
Possible causes and solutions for high background signal:
a. Too much antibody used
-
Reduce the concentration of the primary and/or secondary antibody. Optimize antibody concentrations for your assay
b. Non-specific antibody binding
-
Use a suitable blocking buffer or use an affinity-purified antibody
c. Too much detection reagent used
-
Repeat assay with a higher dilution of detection reagent
d. Too few washing cycles
-
Increase the number of washing cycles
e. Contaminated blocking agent
f. Wrong concentration of blocking agent
-
Check the concentration of blocking agent in the recommended protocol
g. Presence of blocking buffer interferes with antibody binding
-
Wash off blocking buffer before adding antibody
h. Reaction not stopped
-
Use stop solution to prevent overdevelopment
i. Plate left too long before reading
-
Start taking measurements shortly after the addition of detection reagent
j. Wrong settings on the plate reader
-
Check settings and adjust as needed
k. Insufficient amount of Tween® in the buffers
-
Use PBS containing 0.05% Tween®
l. Incubation with substrate carried out in the light
-
Perform substrate incubation in the dark
m. Incubation temperature too high
-
Optimize the incubation temperature for your assay
n. Plates stacked during incubations leading to uneven temperature distribution
o. Pipetting errors
-
Calibrate pipettes so that they dispense the correct volumes
p. Reagents were not mixed properly
-
Before pipetting solutions into wells, make sure all reagents and samples have been throughly mixed
q. Inconsistent washing of wells
-
Take precautions to reduce variability of washes
r. Uneven evaporation of solution from wells during incubation
-
Always incubate with a lid on the plate
s. Interference caused by dirt on the bottom of the plate
-
Clean the plate carefully and reread
Problem 4: Slow color development
Possible causes and solutions for slow color development:
a. Incubation temperature is wrong
-
Ensure plates and reagents are kept at room temperature
b. Contaminated solutions
c. Detection reagent too old, contaminated or used at the wrong pH
-
Use fresh detection reagents at the correct pH